GYIG OpenIR  > 矿床地球化学国家重点实验室
An optimized protocol of high precision measurement of Hg isotopic compositions in samples with low concentrations of Hg using MC-ICP-MS.
Hongyan Geng;  Runsheng Yin;  Xiangdong Li
2018
发表期刊Journal of Analytical Atomic Spectrometry
卷号33期号:11页码:1932-1940
摘要

Multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) enables the precise determination of mercury (Hg) isotopic compositions in different types of samples. However, isotopic measurements in samples with low Hg concentrations and small sample mass remain a challenge. In this study, we developed an optimized protocol for high precision Hg isotope determination using a modified cone arrangement (X skimmer cone + jet cone) with Neptune Plus MC-ICP-MS. Through modification of the HGX-200 continuous-flow cold-vapor generation system, and careful optimization of the instrument gas flow rates, we obtained steady and high signal sensitivity for Hg (202Hg: 1.78 V per ng mL−1). Our method allowed the precise determination of low Hg concentration in solutions (0.10 ng mL−1), which is the lowest according to the literature. Only 0.70 ng Hg in samples was required according to our new analytical method, which enables the direct measurement of low Hg concentration in samples (down to 5 ng g−1), after direct acid digestion treatment. Using our method, the Hg isotopic compositions of four igneous rock standard reference materials (BCR-2, BHVO-2, GSP-2, and GSR-2) were reported for the first time. The standard reference materials showed large variations of δ202Hg (−1.24 to −2.47‰), indicating that mass dependent fractionation (MDF) of Hg isotopes occurred during magmatic processes.

关键词Plasma-mass Spectrometry Mercury Isotopes Preconcentration Sediments Seawater Ratios China
收录类别SCI
语种英语
文献类型期刊论文
条目标识符http://ir.gyig.ac.cn/handle/42920512-1/10403
专题矿床地球化学国家重点实验室
作者单位1.University Research Facility in Chemical and Environmental Analysis (UCEA), The Hong Kong Polytechnic University, Hong Hum, Kowloon, Hong Kong
2.Shandong Provincial Key Laboratory of Depositional Mineralization and Sedimentary Minerals, College of Geological Sciences and Engineering, Shandong University of Science and Technology, Qingdao 266590, China
3.State Key Laboratory of Ore Deposit Geochemistry, Institute of Geochemistry, Chinese Academy of Sciences, Guiyang 550002, China
4.Department of Civil and Environmental Engineering, The Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong
推荐引用方式
GB/T 7714
Hongyan Geng;Runsheng Yin;Xiangdong Li. An optimized protocol of high precision measurement of Hg isotopic compositions in samples with low concentrations of Hg using MC-ICP-MS.[J]. Journal of Analytical Atomic Spectrometry,2018,33(11):1932-1940.
APA Hongyan Geng;Runsheng Yin;Xiangdong Li.(2018).An optimized protocol of high precision measurement of Hg isotopic compositions in samples with low concentrations of Hg using MC-ICP-MS..Journal of Analytical Atomic Spectrometry,33(11),1932-1940.
MLA Hongyan Geng;Runsheng Yin;Xiangdong Li."An optimized protocol of high precision measurement of Hg isotopic compositions in samples with low concentrations of Hg using MC-ICP-MS.".Journal of Analytical Atomic Spectrometry 33.11(2018):1932-1940.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
An optimized protoco(563KB)期刊论文作者接受稿开放获取CC BY-NC-SA浏览 请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Hongyan Geng;Runsheng Yin;Xiangdong Li]的文章
百度学术
百度学术中相似的文章
[Hongyan Geng;Runsheng Yin;Xiangdong Li]的文章
必应学术
必应学术中相似的文章
[Hongyan Geng;Runsheng Yin;Xiangdong Li]的文章
相关权益政策
暂无数据
收藏/分享
文件名: An optimized protocol for high precision measurement of Hg isotopic compositions.pdf
格式: Adobe PDF
此文件暂不支持浏览
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。