GYIG OpenIR  > 矿床地球化学国家重点实验室
Two reliable calibration methods for accurate in situ U-Pb dating of scheelite
Yanwen Tang; Junjie Han; Tingguang Lan; Jianfeng Gao; Liang Liu; Changhao Xiao; Jiehua Yang
2022
发表期刊Journal of Analytical Atomic Spectrometry
卷号37期号:2页码:358-368
摘要

Scheelite is an important metallic mineral in W-related hydrothermal deposits and can be utilized as a reliable geochronometer to directly date the timing of mineralization. Up to now, two previous studies have performed in situ U–Pb dating of scheelite using NIST glasses as reference materials due to the lack of scheelite standards. However, there exists a significant matrix effect between NIST612 and scheelite based on our analysis. Therefore, two reliable calibration methods have been proposed and assessed via LA-SF-ICP-MS in this study. A laser spot size of 32 μm was mainly used to satisfy the signal for 206Pb, 207Pb, and 238U and keep a high resolution to fit small grains and altered crystal zones of scheelite samples. Scheelite WX27 as a secondary standard has been accurately determined in normal ablation settings of 3 J cm−2 – 5 Hz, 3 J cm−2 – 10 Hz, 5 J cm−2 – 5 Hz (using a spot size of 24 μm), 5 J cm−2 – 10 Hz, and 10 J cm−2 – 5 Hz. Its lower intercept 206Pb/238U ages are consistent with each other within error and agree with two known ages (144.8 ± 11.7 and 141.8 ± 5.3 Ma) from the Woxi deposit with age offsets of <3.8%, no matter whether they were calibrated by using concordant wolframite YGX or the combined calibration method (i.e., NIST612 for 207Pb/206Pb and YGX for 238U/206Pb ratios). Moreover, accurate U–Pb ages, ranging from 432 Ma to 92 Ma, were also obtained for five typical W deposits. All these age results demonstrate the reliability of the two calibration methods and indicate that no matrix effect exists between scheelite and wolframite using single spot analysis mode in normal ablation settings. When all the same data from scheelite WX27 were calibrated by using NIST 612, most of the obtained ages were much younger than the known ages, with the age offsets ranging from 5.9% to 18.0%. Thus, NIST glasses are commonly not suitable as primary standards for in situ U–Pb dating of scheelite. Scheelite WX27 contains relatively high U (averaging 7.1 ppm) and low common-lead and is suitable as a candidate reference material for in situ U–Pb dating of scheelite at the moment.

DOI10.1039/D1JA00387A
URL查看原文
收录类别SCI
语种英语
引用统计
被引频次:15[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.gyig.ac.cn/handle/42920512-1/13631
专题矿床地球化学国家重点实验室
作者单位1.State Key Laboratory of Ore Deposit Geochemistry, Institute of Geochemistry, Chinese Academy of Sciences, Guiyang, China
2.Laboratory of Dynamic Diagenesis and Metallogenesis, Institute of Geomechanics, Chinese Academy of Geological Sciences, Beijing 100081, China
推荐引用方式
GB/T 7714
Yanwen Tang,Junjie Han,Tingguang Lan,et al. Two reliable calibration methods for accurate in situ U-Pb dating of scheelite[J]. Journal of Analytical Atomic Spectrometry,2022,37(2):358-368.
APA Yanwen Tang.,Junjie Han.,Tingguang Lan.,Jianfeng Gao.,Liang Liu.,...&Jiehua Yang.(2022).Two reliable calibration methods for accurate in situ U-Pb dating of scheelite.Journal of Analytical Atomic Spectrometry,37(2),358-368.
MLA Yanwen Tang,et al."Two reliable calibration methods for accurate in situ U-Pb dating of scheelite".Journal of Analytical Atomic Spectrometry 37.2(2022):358-368.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Two reliable calibra(7848KB)期刊论文作者接受稿开放获取CC BY-NC-SA浏览 请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Yanwen Tang]的文章
[Junjie Han]的文章
[Tingguang Lan]的文章
百度学术
百度学术中相似的文章
[Yanwen Tang]的文章
[Junjie Han]的文章
[Tingguang Lan]的文章
必应学术
必应学术中相似的文章
[Yanwen Tang]的文章
[Junjie Han]的文章
[Tingguang Lan]的文章
相关权益政策
暂无数据
收藏/分享
文件名: Two reliable calibration methods for accurate in situ U-Pb dating of scheelite.pdf
格式: Adobe PDF
此文件暂不支持浏览
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。