A novel experimental device for electrochemical measurements in supercritical fluids up to 700 °C/1000 bar and its application in the corrosion study of superalloy Inconel 740H
Sen Lin;  Heping Li;  Liping Xu;  Yanqing Zhang;  Can Cui
2017
Source PublicationRSC Advances
Issue7Pages:33914-33920
Abstract

A three-electrode electrochemical study concerning electrically conductive materials submitted to supercritical fluids above 550 °C has not been reported in spite of its significance in many related fields. In this work, the design of a novel experimental device allowing three-electrode electrochemical measurements in high temperature and pressure fluids up to 700 °C/1000 bar is detailed described. In this device, pyrophyllite is used as a new electrode sealing material and the electrodes are sealed using a conical self-energizing sealing structure, which ensures the water tightness of the electrodes at pressures up to 1000 bar. The reference electrode in this device can either be an Ag/AgCl external pressure balanced reference electrode (EPBRE) or a platinum reference electrode. The potentiodynamic polarization plots and electrochemical impedance spectroscopy (EIS) of nickel-based superalloy Inconel 740H, a candidate material for Generation IV nuclear reactors, in deaerated water and chloride containing fluid (600–700 °C, 600–1000 bar) are presented using this device.

Language英语
Document Type期刊论文
Identifierhttp://ir.gyig.ac.cn/handle/42920512-1/8360
Collection地球内部物质高温高压实验室
Affiliation1.Key Laboratory for High Temperature & High Pressure Study of the Earth's Interior, Institute of Geochemistry, Chinese Academy of Sciences, Lincheng West Road 99, Guiyang 550081, China
2.National Laboratory for Clean Energy (DNL), Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Zhongshan Road 457, Dalian 116000, China
Recommended Citation
GB/T 7714
Sen Lin;Heping Li;Liping Xu;Yanqing Zhang;Can Cui. A novel experimental device for electrochemical measurements in supercritical fluids up to 700 °C/1000 bar and its application in the corrosion study of superalloy Inconel 740H[J]. RSC Advances,2017(7):33914-33920.
APA Sen Lin;Heping Li;Liping Xu;Yanqing Zhang;Can Cui.(2017).A novel experimental device for electrochemical measurements in supercritical fluids up to 700 °C/1000 bar and its application in the corrosion study of superalloy Inconel 740H.RSC Advances(7),33914-33920.
MLA Sen Lin;Heping Li;Liping Xu;Yanqing Zhang;Can Cui."A novel experimental device for electrochemical measurements in supercritical fluids up to 700 °C/1000 bar and its application in the corrosion study of superalloy Inconel 740H".RSC Advances .7(2017):33914-33920.
Files in This Item:
File Name/Size DocType Version Access License
A novel experimental(1076KB)期刊论文作者接受稿开放获取CC BY-NC-SAView Application Full Text
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[Sen Lin;Heping Li;Liping Xu;Yanqing Zhang;Can Cui]'s Articles
Baidu academic
Similar articles in Baidu academic
[Sen Lin;Heping Li;Liping Xu;Yanqing Zhang;Can Cui]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[Sen Lin;Heping Li;Liping Xu;Yanqing Zhang;Can Cui]'s Articles
Terms of Use
No data!
Social Bookmark/Share
File name: A novel experimental device for electrochemical.pdf
Format: Adobe PDF
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.